Sign In | Join Free | My fnxradio.com
China China World Technology Medical Equipment Service Group logo
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group Win-win together!
Verified Supplier

5 Years

Home > Patient Monitor Parts >

Hospital Medical Parts Patient Monitor ECG Board PHiliph Goldway UT6000A C-ECG7131C v1

China World Technology Medical Equipment Service Group
Trust Seal
Verified Supplier
Credit Check
Supplier Assessment
Contact Now

Hospital Medical Parts Patient Monitor ECG Board PHiliph Goldway UT6000A C-ECG7131C v1

Brand Name : Goldway

Model Number : UT6000A

MOQ : 1

Payment Terms : T/T, Western Union,PayPal

Delivery Time : 3-5 working days

Product name : Goldway ECG Board

Type : UT6000A

Application : medical equipment,ICU,hospital device

Warranty : 90days

PN : C-ECG7131C v1

Material : Plastic, PCB

Apply For : Patient Monitor

Product Edge : Technical engineers with 12 years of experience

Post-sales service : If we find non-human damage, we will be free for you to repair or returned

Supported language : English,French,Italian,Spanish,Japanese,Hindi,Portuguese,Russian,Arabic,etc

Product State : In Good Condition

Product Peculiarity : Professional, Reliable,Economic

MOQ : 1 pc

HR measurement range : 15 – 300 bpm

Accuracy : ±0.02mV or ±5%

ST measurement range : ECG 1 only; -0.8mV – +0.8 mV

Sweep speed : 12.5 mm/s, 25 mm/s, 50 mm/s

Contact Now

​​ PHiliphh Goldway UT6000A Patient Monitor ECG Board C-ECG7131C v1

Hospital Medical Parts PHiliphh Goldway UT6000A Patient Monitor ECG Board C-ECG7131C v1


► Brief Introduction

The Goldway UT6000A ECG Board is a component of the Goldway UT6000A series electrocardiograph (ECG) machines. It is responsible for acquiring and processing the electrical signals from the electrodes placed on a patient's body to generate an electrocardiogram. The ECG board converts the analog electrical signals into digital data, filters out noise, and performs various calculations to display and interpret the ECG waveform. It is a crucial part of the ECG machine that helps in diagnosing heart conditions and monitoring cardiac activity.

► Performing ECG Test

To set the ECG Simulator:

  • Set the HR to 80bpm and ECG gain to 1mV.

To test the ECG:

  • Connect one end of the ECG cable to the ECG connector of the monitor and the other end to the simulator.
  • Confirm that the patient simulator has been properly connected to the monitor.
  • Confirm and make sure the ECG lead II is displayed on the ECG channel, without noise and the heart rate displays 80±2bpm. When the R wave appears you should hear a beep.
  • Confirm every ECG lead and make sure there is an ECG waveform and no noise interference.
  • Adjusting the ECG gain to 2mV, and confirm the change in the ECG waveform.
  • Unplug the RA from the simulator, and confirm that a lead off message pops-up.
  • Reconnect RA lead.
  • Set the 1mV calibration in the ECG menu, and confirm that a 1mV calibration square wave is overlapped on the ECG waveform.
  • The ECG test is complete

► Removing the ECG Module

  1. Disassemble the main frame.
  2. Remove the screws. (Circles)
  3. Disconnect the ECG connector. (Rectangle)
  4. Remove the ECG module.

Hospital Medical Parts Patient Monitor ECG Board PHiliph Goldway UT6000A C-ECG7131C v1

Specification

Lead type: Standard 3-Lead or 5-Lead selectable
Lead selection: I, II, III or I, II, III, aVR, aVL, aVF, V
Display channels: 2 channels, ECG1, ECG2
HR measurement range: 15 – 300 bpm
Accuracy: ±5 bpm or ±1%, whichever is greater
ST measurement range: ECG 1 only; -0.8mV – +0.8 mV
Accuracy: ±0.02mV or ±5%
Sweep speed: 12.5 mm/s, 25 mm/s, 50 mm/s
Gain selection: x1/4, x1/2, x1, x2, x4, Auto
Pacemaker detection: Yes
Protection: Against electrosurgical interference and defibrillation

► More Services

In addition to the sales of medical equipment, accessories and maintenance accessories, we also specialize in the maintenance of such machines and modules. As we are proficient in the maintenance of hospital equipment, we are good at renovating damaged machines and accessories. Therefore, if you have such machines or modules for maintenance, you can contact us and send them to us, and we will repair them at the fastest speed. But it is important to note that you must make sure that the items to be serviced are not opened and never serviced. Otherwise, it is unacceptable to us. Finally, please check your items carefully before sending them out.

Looking forward to receiving your early inquiry about PHiliphh Goldway UT6000A Patient Monitor ECG Board C-ECG7131C v1. Thank you for your time!


Product Tags:

UT6000A Patient Monitor ECG Board

      

C-ECG7131C v1 Patient Monitor ECG Board

      
Wholesale Hospital Medical Parts Patient Monitor ECG Board PHiliph Goldway UT6000A C-ECG7131C v1 from china suppliers

Hospital Medical Parts Patient Monitor ECG Board PHiliph Goldway UT6000A C-ECG7131C v1 Images

Inquiry Cart 0
Send your message to this supplier
 
*From:
*To: China World Technology Medical Equipment Service Group
*Subject:
*Message:
Characters Remaining: (0/3000)